HE-960RW

面板安装型双通道电阻率计(电阻率计)

用於控制和监测纯水和超纯水的处理和生产过程。

透过採用新设计的温度测量电路,高阶电阻率表可显着改善鲍笔奥(超纯水)和高温纯水的电阻率测量的关键温度补偿性能。高度稳定的电阻率测量是透过不仅在水温下而且在环境温度下的变化来实现的。

RESIST

产物分类: Process Water
製造商: 蓝鲸体育直播 Advanced Techno Co., Ltd.

Enables measurements with minimal error

Measurements are affected less by ambient temperature and sample temperature, enabling accurate measurements with less error.

Highly responsive temperature sensor

Helps to minimize fluctuations in measurement values due to delayed temperature compensation in the event of sudden changes in water temperature.

We provide equipment and sensor calibration services

Temperature sensor offset calibration is offered as well as transmitter and sensor combination calibration services. *Option

2-channel specification

Allowing measurement at 2 points, and capable of displaying salt removal rate.

Product nameIndustrial 2-channel specific resistance meter
ModelHE-960RW
Measurement
method
Electrode type (2-electrode method)
Sensor input2 channels (for concurrent measurement with sensors isolated each other)
Cell constant0.01/cm
Temperature sensor
specifications
Platinum resistance 1000 Ω/0℃
Measuring rangeMeasurement unit办Ω?尘0 to 20.00 to 200.00 to 1000 *
惭Ω?肠尘0 to 2.000 to 20.000 to 100.0 *
*: Measurable without temperature compensation
Temperature????
??
0℃ to 100℃
(The number of digits after the decimal point is selectable from one and two. The range of measurements is ?20℃ to 120℃.)
Desalination rate0% to 100%
Temperature: 0℃ to 100℃ (Select your desired decimal point from 0, 1, and 2 digits.)
RepeatabilityWithin ±0.1% of the full scale (in equivalent input)
LinearityWithin ±0.5% of the full scale (in equivalent input)
Transmission
output
Number of outputs: 2
4 mA to 20 mA DC/0 mA to 20 mA DC, input/output isolated type
Maximum load resistance: 900 Ω
Transmission output range: Freely selectable within the measurement range
(Negative terminals of each transmission output channel are connected inside and thus have the same electric potential.)
Contact outputNumber of outputs: 4
Alarm contact outputs (R1, R2, and R3)
  Contact type: relay contact, SPST (1a)
  Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load)
  Contact function: selectable from the upper/lower limit action (ON/OFF control), delay, and hysteresis
  Selectable from the selected measurement, anomaly alarm, and maintenance.
Alarm contact output (R4)
  Contact type: relay contact, SPDT (1c)
  Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load)
  Contact function: selectable from the upper/lower limit action (ON/OFF control), delay, and hysteresis
  Selectable from the selected measurement, anomaly alarm, and maintenance.
  (However, R1 and R2, and R3 and R4 share the common contacts respectively.)
Calibration functionSpecific resistance: Based on the specified compensation coefficient for cell constant (parameter input)
Temperature: Calibrated by comparing with the reference thermometer
Communication
output
RS-485 input/output
Transmission
output hold feature
  • Previous value hold
  • Optional value hold
    Selectable from the above (However, only the previous value hold is available in the maintenance mode.)
Self-diagnosis
function
  • Sensor diagnosis
    Short-circuit and disconnection of the temperature sensor
  • Out of the measurement range
  • Converter error
Temperature
compensation
  • Temperature compensation for extra-deionized water and impurities
      Select the temperature characteristics of impurities
      - NaCl temperature characteristics
      - Arbitrary temperature coefficient entry (temperature coefficient: 5%/℃)
      Reference temperature: 5℃ to 95℃
  • No temperature compensation is performed.
Temperaturecompensated
range
0℃ to 100℃
Extra deionized
water specific
resistance selection
Measurement
unit
办Ω?尘182.3 (standard), 181.8, 182.4Select from options shown on the left.
惭Ω?肠尘18.23 (standard), 18.18, 18.24
Clipping functionWhen the measured value is above the upper limit of the measurement range derived from the specified specific resistance, the specified resistance is used as the measured value.
Ambient
temperature
?5℃ to 45℃
Relative humidity20% to 85% (without dew condensation)
Storage
temperature
?25℃ to 65℃
Power supplyRated voltage 100 V to 240 V AC, 50/60 Hz, 10 VA (max.)
StructureIndoor-use panel installation type
Panel case: ABS, Terminal: PBT
Panel: IP65 dust and water proof structure
Protective structurePanel: IP65 (IEC60529, JIS C0920)
Rear case: IP20, Terminal:IP00
Class II device (IEC61010-1)
Pollution level 2 (IEC61010-1)
Conforming
standards
CE MarkingEMC Directives: EN61326-2-3
Low Voltage Directive: EN61010-1
RoHS Directive: EN50581
FCC RuleFCC Part15
External
dimensions
96 (W) mm × 96 (H) mm × 115 (D) mm
Case depth: approx. 105 mm (when panel-mounted)
MassApprox. 550 g
Compatible sensorsERD/ERF-series specific resistance sensor, cell constant 0.01/cm