Available for US OEM customers only
Spectral coverage | With 532 nm laser:210 to 3500 cm鈥1 With 660 nm laser:400 to 3150 cm鈥1 With 785 nm laser:210 to 2700 cm鈥1 |
Numerical aperture | F/3.1 |
Stray light rejection Typical (Maximum) | 0.1% (0.2%) (measured with bandpass filter 780BP10) |
CCD Detector & Typical Q.E. | Back-illuminated CCD with low etaloning in NIR 55% peak QE at 850nm(VIS optimized) |
Detector height Fiber-optic option | 300渭m CCD height standard (1000渭m optional) 200 or 300渭m dia.,1m long fiber optic (for 1mm tall CCD, RTS bundle is offered) |
Thermoelectric stabilization | None. Dark current and CCD-pattern noise must be subtracted. User must shutter in optical path. QE shifts slightly with temperature. |
Spectral resolution Pixel resolution Slit (factory configuration) | 785nm configuration;25渭m slit; 2048 pixels; 8鈥12cm鈥1 resolution; 0.19nm/pixel (configuration with 300渭m tall CCD) |
Improved CCD full well Raw non linearity Factory corrected non linearity | >250Ke- (sensitivity mode) < 0.4% (sensitivity mode) |
Typical dynamic range | 7000:1 in sensitivity mode |
Readout speed | 8.6ms (500kHz mode); 116 spectra/s with 0 exposure time (Multi-Acq mode) Max: 4.5 ms (Ultra mode); 223 spectra/s with 0 exposure time (Multi-Acq mode) |
Typical Dark current | 1 count/ms at 20掳C (room temp.); typical offset = 1000 counts |
Typical readout noise | 35 e- (Max= 45e-) in sensitivity mode (4 e鈥/count) |
A/D converter | 16 bit, 500 kHz (pixel rate); optional 2 MHz VIS-CCD |
Typical dynamic range | 7,000:1 in sensitivity mode |
Weight | 1.8 lb (0.82 kg) |
Gain selection | 4 e-/count and 8 e-/count |
Dimensions (H脳 W脳D) | 4鈥 脳 2.71鈥 脳 1.49鈥 (101.6 mm 脳 68.8 mm 脳 37.8 mm) |
* Specifications, Form Factor and Spectrometer Covers are subject to change without notice. |
Acquisition software included (LabVIEW鈩 2011 only)
- VIs and top-level code are provided for customer customizations
- Access to data with raw CCD linearity and corrected linearity done at factory for each CCD chip
- CCD settings and dark-subtract
- On-board or software averaging
- Scale selection between pixel, wavelength, and wavenumber
- On-board spectral calibration
- Linearity correction on/off
- save function to Excel庐 or text 铿乴e
No LabVIEW鈩 license is needed to run our acquisition software. LabVIEW鈩 license ver. 2011 required to edit our code. No code customization supported in price.